Roche CoaguChek XS System Manual do Operador Página 104

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 103
Control Testing and Quality Control
104
Performing a liquid quality control test
1 Place the meter on a level, vibration-free
surface or hold it in your hand so it is
roughly horizontal.
2 Power the meter on by pressing the
button for approximately 1 second.
You can also power on the meter directly by
inserting a test strip or connecting the power
adapter.
3 Wait until the Main Menu is displayed, or
log on as described on page 86.
Main Menu 09:15 am
04/19/2012
Control Test
Review Results
Setup
Patient Test
Vista de página 103
1 2 ... 99 100 101 102 103 104 105 106 107 108 109 ... 155 156

Comentários a estes Manuais

Sem comentários