Roche Accu-Chek Active Manual do Utilizador Página 44

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 43
84
Hold the test strip so that the application area and arrows are
facing upwards. Gently push the test strip in the direction of
the arrows into the test strip guide
of your Accu-Chek Active
monitor, until you hear it click into place.
Inserting the test strip automatically puts the monitor in Test Mode.
Please remember that your Accu-Chek Active monitor automatically
turns off after about 1–2 minutes of non-use (i.e. when no button
is pressed). If this happens, remove the test strip and repeat the
procedure described above with a new test strip.
Now watch the display.
85
Remove a test strip from its container. Close the container
immediately. The cap contains a drying agent which ceases to
function if the container is left open, rendering the test strips
unusable.
Check the round control window on the back of the test strip
against the colour scale printed on the test strip container.
The colour of the control window must match that of the
colour interval at the top (0 mg/dL / 0 mmol/L).
If the test strip shows a different colour, do not use it.
6.3 Quality control test procedure
015134 AccuChek Active_Eng.indd 84-85 12.08.2004, 14:54:36