Wipe the measurement optics components with a soft lint-
free cloth and/or a cotton swab. The cloth/cotton swab may
be slightly moistened with water or 70% alcohol. Make sure
that no liquid enters the monitor itself. Avoid scratching the
measurement optics, as this will impair the measuring function.
When all of the components are thoroughly dry, you may slide
the test strip guide back onto the monitor. Ensure it clicks back
into place.
Then perform a quality control test (see Chapter 6, “Checking
Accu-Chek Active”).