Roche Accu-Chek Aviva Informações Técnicas Página 232

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 231
230
Display Shows Possible Cause(s) Possible Solution(s)
Above Hyper
Warning Limit
j Your test result is above the
hyper warning limit set in the
meter.
j Treat your high blood glucose as recommended by
your healthcare professional.
j Consider checking bG, ketones, and insulin.
j See the pump User Guide for additional solutions.
Beep/Vibrate O j Beep and Vibrate settings
cannot be turned o at the
same time.
j Ensure that either the Beep or Vibrate setting is
turned on. See Chapter 5, “Changing Meter
Settings.
Below Hypo
Warning Limit
j Your test result is below the
hypo warning limit set in the
meter.
j Treat your low blood glucose as recommended by
your healthcare professional.
j The meter displays a recommended number of
carbohydrates for you to eat, and then retest your
blood glucose.
Bolus Advice
Bolus Advice data
deleted
j Bolus advice data integrity
cannot be confirmed.
j Wait 8 hours for accurate bolus advice.
IMPORTANT:
Insulin doses and meals taken before the Bolus Advice warning are no longer
reflected in the bolus advice calculation.
The meter may not reflect the pump bolus history, however the bolus history is
available on the pump.
50813_00048009694_D.indb 230 7/3/12 3:13 PM
Vista de página 231
1 2 ... 227 228 229 230 231 232 233 234 235 236 237 ... 281 282

Comentários a estes Manuais

Sem comentários