Roche Accu-Chek Aviva Informações Técnicas Página 236

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 235
234
Display Shows Possible Cause(s) Possible Solution(s)
Connection Lost j The connection was lost
between the meter and the
pump during the pairing
process. Therefore, the
attempt to pair the meter and
pump was unsuccessful.
j Restart the pairing process. See the Advanced
Owners Booklet.
j Pump is out of range of the
meter.
j Ensure the meter and pump have Bluetooth wireless
technology turned on.
j Ensure the meter and pump are within
communication range.
j If the low battery icon is displayed on the meter, the
meter cannot communicate with the pump. Replace
the batteries.
Diary Entry Used
For Advice
j The diary entry selected has
been used for bolus advice
and modifications are not
allowed.
j Carefully confirm all information involving bolus
advice. You cannot modify bolus advice related
entries on the meter.
Diary Results
Expired
j Bolus advice is no longer
available for this result.
j Review data that was saved in My Data. Then,
decide to modify data, add data, or begin a new
bolus advice session.
50813_00048009694_D.indb 234 7/3/12 3:13 PM
Vista de página 235
1 2 ... 231 232 233 234 235 236 237 238 239 240 241 ... 281 282

Comentários a estes Manuais

Sem comentários