232
Display Shows Possible Cause(s) Possible Solution(s)
Bolus Too High j The bolus is above the
allowed amount. The bolus
was set to 50 Units.
j Check the accuracy of all entries. If necessary,
contact your healthcare professional.
Bolus Too Low j The insulin pump cannot
deliver a 0.1 Unit of multiwave
bolus.
j The pump will adjust the bolus to 0.2 Units.
Carbohydrate Ratio j Carbohydrate ratio is outside
of the acceptable meter
range.
j Check your entries and contact your healthcare
professional to determine the appropriate settings.
Code Key Missing j The activation chip is not
inserted.
j Turn the meter o and insert the activation chip. If
you need an activation chip, please contact Roche.
Communication
Lost
j Pump is out of range of the
meter.
j Bolus delivery continuing, see pump screen.
j Ensure the meter and pump have Bluetooth wireless
technology turned on.
j Ensure the meter and pump are within
communication range.
j If the low battery icon is displayed on the meter, the
meter cannot communicate with the pump. Replace
the batteries.
j Use the pump to monitor or cancel a bolus which is
in the process of being delivered.
50813_00048009694_D.indb 232 7/3/12 3:13 PM