Preparing to run a liquid quality control test
Prepare for a liquid quality control test in the same way you would prepare to perform a
test with a capillary blood sample. The only difference is the use of control solution instead
of blood.
Quality control
70
1 Have the test strip container at hand.
2 If you are using the test strip lot for the
first time, make sure that the test strip
code chip that came with these test
strips is at hand.
3 Make sure the bottle of freeze-dried
control plasma and the dropper for
making the control solution are at hand.
This bottle should remain refrigerated
(not frozen) until use.
4 Make sure that the quality control code
chip that came with the control solution
is at hand.
5 Open the lid of the bottle and remove
the rubber cap.
6 Hold the dropper with the sealed
dropper neck pointing upward, then cut
off the end of the cap with scissors. Do
not hold the dropper close to your face.
Caution: To avoid loss of diluent, hold the
dropper by the stem; do not squeeze the
bulb of the dropper while cutting the tip.
Apply gentle pressure to the reservoir
to transfer the entire contents of the
dropper to the bottle. Make sure that the
dropper does not come in contact with
the dried control plasma.