Quality control
75
QC Test
09:15 am
09/16/2009
QC: 999
Code: 184
Op: Operator 1
Par: PT
Level 1
Level 2
QC Test
09:15 am
09/16/2009
QC: 999
Code: 184
Op: Operator 1
Par: PT
Level: 1
QC Test
09:15 am
09/16/2009
180 SEC
QC: 999
Code: 184
Op: Operator 1
Par: PT
Level: 1
20 Select the level for this measurement.
The hourglass symbol shows that the test
strip is warming up. When the warming-
up process is complete, a further beep
tone (provided the beeper is turned on)
indicates that you can now apply control
solution.
The dropper symbol flashes to indicate
that the meter is ready to perform the
test and is waiting for the sample to be
applied.
At the same time, a 180-second
countdown begins. You must apply the
sample within this time; otherwise, you
will receive an error message.