Introduction
19
Note on systems with older
hardware
The Accu-Chek Inform II system complies with the emis-
sion and immunity requirements described in IEC 61326-
2-6:2006. It has been designed and tested to CISPR 11
Class A. In a domestic environment it may cause radio
interference, in which case, you may need to take meas-
ures to mitigate the interference. The electromagnetic
environment in which the Accu-Chek Inform II device is
to be operated should be thoroughly evaluated prior to
operation of the device.
WARNING
Possibility of Electromagnetic Interference
Under certain conditions, a specific combination of hard-
ware equipment may emit electromagnetic radiation of
such high energy that it can interfere with the proper
functioning of other electronic or medical equipment.
Therefore it is not allowed to dock an Accu-Chek Inform
II meter with a serial number above or equal to
UU11030000, which is equipped with an RF card, into an
Accu-Chek Inform II Base Unit with a serial number
below UU41030000 if this Accu-Chek Inform II Base Unit
is connected to a network directly or through a PC via a
cable.
Note: The following classification applies if an
Accu-Chek Inform II meter (with RF card) with a serial
number below UU11030000 is docked in an
Accu-Chek Inform II Base Unit with a serial number
below UU41030000 if this Accu-Chek Inform II Base
Unit is connected to a network directly or through a
PC via a cable.