Roche ACCU-CHEK Inform II Manual do Operador Página 55

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 54
Glucose Control Testing
55
Applying the control solution Once the meter has detected the test strip, you are
prompted to apply the control solution.
1 Wait until the flashing drop appears in the display
before applying the control solution.
2 Apply a drop of glucose control solution to the
front edge of the test strip. Do not apply the
control solution to the top of the strip.
The control solution is pulled into the test strip by
capillary action.
Once sufficient control solution has been detected, the
meter beeps and the measurement begins.
Control Test 12:48pm
03/16/10
Control L2 (Hi) 123456
Strip Lot 545794
Vista de página 54