Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
67
Storing control solution information
Glucose control solution lot information can be entered
before testing, if lot editing has been allowed at the meter
level in the setup, and appears in a list for operators to
refer to. Use the following procedure to add glucose con-
trol lot numbers to the Control Lot list.
Entering the lot number of the
control solution
The following description assumes that the meter is pow-
ered on and the main menu is displayed.
1 Touch to open the Main Menu 2 screen.
2 Touch Control Lots to open the related menu.
3 Touch Add to enter a new lot number.
Main Menu 12:48pm
03/16/10
Review Results
Control Test
Patient Test
Main Menu 2 12:48pm
03/16/10
Maintenance
Strip Lots
Proficiency
Control Lots
Beeper
Linearity
Diagnostics
Linearity Lots
Admin.
Control Lots
03/16/10
12:48pm
Add
*QC L2 03/15/11 777678
12/31/10 777732
03/15/11 777723
QC L1
QC L2
Type Exp. Date Lot Num.