Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
62
Transferring code key
information to the meter
The following description assumes that the meter is pow-
ered on and the main menu is displayed.
1 Touch to open the Main Menu 2 screen.
2 Touch Strip Lots to open the related menu.
3 Touch Add if you want to add the information for a
new test strip lot from a new code key. The Add
Strip Lot screen opens.
4 Insert the new code key in the opening of the code
key reader. An LED starts flashing and signals that
the reader is ready to transfer data.
Main Menu 12:48pm
03/16/10
Review Results
Control Test
Patient Test
Main Menu 2 12:48pm
03/16/10
Maintenance
Strip Lots
Prociency
Control Lots
Beeper
Linearity
Diagnostics
Linearity Lots
Admin.
Strip Lots
03/16/10
12:48pm
Add
Strip
12/31/10 855732
* 03/15/11 845678
Strip 03/15/11 845723
Type Exp. Date Lot Num.
Strip
The code key reader remains ready to send for a few
seconds after it has transmitted data. As a result, you
can perform the following procedure on multiple
meters without having to reinsert the code key.