Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
68
4 Select the level (L1/Lo or L2/Hi).
5 Use the keypad to enter the lot number. Touch
to confirm the entered lot number, or
6 Touch to read the lot number from the
control solution bottle via barcode scanner. Follow
the instructions for scanning IDs (see page 31). *
7 Use the key pad to enter the expiration date as
stated on the control solution bottle.
8 Touch to confirm the entered expiration date.
* Barcodes on test strips vials/control solutions may not
be available in all countries (see page 39).
123
456
789
0
Control Lot 12:48
pm
03/16/10
Add Control Lot 12:48pm
03/16/10
Level 2 (Hi)
Level 1 (Lo)
123
456
789
0
Control Expiration 12:48
pm
03/16/10
07/31/10