Initial Startup
94
Installing or replacing the battery pack
When shipped, the battery pack is not installed in the
Accu-Chek Inform II meter.
After installing a new battery pack, the meter should be
charged for three hours in the base unit before testing.
Whenever the meter is in the base unit, the icon
is displayed. This icon shows that power is available
and the meter can charge, if necessary.
Make sure that the allowable temperature range for
charging the battery pack (3-40 °C or 37-104 °F) is
maintained during installation and initial setup.
When installing a new battery pack in a meter that has
been in use, we recommend downloading the meter
prior to removing the battery pack. This prevents any
inadvertent loss of data.