Initial Startup
115
Value ranges (normal, critical, reportable)
This menu allows you to set limit values for results as
Normal, Critical, or Reportable. Results outside these
limits will be agged to alert you of this event. The
Reportable Range allows the system administrator to set
an institution dened range for reporting patient results.
Values for range limits are entered one directly after the
other in the following order:
n Normal (low/high limit value)
n Critical (low/high limit value)
n Reportable (low/high limit value)
1 Use the keypad to enter the rst value.
2 Touch
to enter the following value, or
3 Touch
to return to the previous view.
4 After entering the last value, touch
to save all
the previous settings.