Linearity Testing
83
Performing a linearity test
Overview of test procedure
A linearity test comprises the following steps:
n Check the lot number of the linearity solutions.
n Check the lot number of the test strips.
n Perform the test with a minimum of three linearity
solutions.
Starting a linearity test
The following description assumes that the meter is
powered on and the main menu is displayed.
1 Touch in the Main Menu to open the Main
Menu 2 screen.
2 Touch Linearity to start the linearity test. The
Linearity Test screen opens.