Linearity Testing
82
Information stored during
linearity testing
The following information is stored for every linearity test:
n Test result
n Lot number of the linearity solution
n Level of linearity solution (L1 to L6)
n Operator ID (if congured)
n Lot number of the test strips
n Time and date of test
n Comments (if applicable)
Linearity test kit
The linearity test kit contains glucose solutions in six
levels (6 vials, 2.5 mL each). For additional information
about the contents and handling of the kit, refer to the
package insert.
Preparing to run a linearity test
Aside from special preparations (see the following
sec tion), a linearity test is run the same as a patient test.
Please check the following:
n At least one code le for test strips must be stored
in the meter and match the lot number of the test
strips used (see Chapter 6).
n The proper test strips must be available.
n An operator ID must be entered (with password, if
applicable), if the meter is congured for login.