Supplement for Other Test Entry
175
11 Select the result of the Control bar test as
acceptable or not acceptable by touching the
appropriate button.
12 Select the result of the liquid control test (Ctrl
Result) as acceptable or not acceptable by
touching the appropriate button.
Once the results are entered, the Pregnancy Control result
screen is displayed.
13 To change an entry or a result, touch the
appropriate button.
To add comments (refer to Adding Comments in this
manual for more information):
14 In the result screen, touch
.
15 Select the desired predened comment(s) from the
display list (if congured) or
16 Touch to enter your own custom comment(s).
Use the keypad (as with login) to enter your
comment.
17 Once you have selected the desired comment(s),
touch
to return to the results screen.
18 Touch
to return to the Main Menu.
Test results are also saved when the meter is powered
off in the results screen or automatically powers off.