Roche ACCU-CHEK Inform II Manual do Operador Página 49

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 48
Patient Glucose Testing
49
The following messages may appear in addition to the
standard test result:
n CR LO (below the Critical Range threshold, but
within the Reportable Range)
n CR HI (above the Critical Range threshold, but
within the Reportable Range)
n RR LO (below the Reportable Range threshold, but
within the System Measurement Range)
n RR HI (above the Reportable Range threshold, but
within the System Measurement Range)
n LO (below the System Measurement Range)
n HI (above the System Measurement Range)
For test results that lie outside the critical or reportable
range, a message (up to 100 characters in length) can
be congured during setup. This message is then
displayed with the corresponding test results.
Instructions on how to add comments are provided in the
following section.
If comments are not set as required and you do not wish
to add a comment to the test result, touch
to return
to the Main Menu.
Test results are also saved when the meter is powered
off in the results screen or when it powers off
automatically.
Remove the test strip and dispose of it in accordance with
applicable regulations and directives for disposal of
potentially infectious samples and materials.